EC calibration solution for Aquamaster meters that specify this solution.
NIST Standard calibration solution.
The EC value depends on the temperature of the solution - for example, the solution has value of 1413 at 25°C
This solution is temperature dependant. See below for details.
| Temperature | Value |
| 15°C | 1147 |
| 20°C | 1278 |
| 21°C | 1305 |
| 22°C | 1332 |
| 23°C | 1359 |
| 24°C | 1386 |
| 25°C | 1413 |
| 26°C | 1440 |
| 27°C | 1467 |
| 28°C | 1494 |
| 29°C | 1521 |
| 30°C | 1548 |